Pappas, Michael (Committee chair)
Herman, Harry (Committee member)
Koplik, Bernard (Committee member)
Date:
1983
Keywords:
Biomechanics--Research
Orthopedic implants
Femur
Availability:
Unrestricted
Abstract:
Femoral component loosening and Prosthetic stem fractures have renewed interest in conservative alternatives to conventional total hip replacements. One such alternative is the concept of surface replacement. The objective in this study is a finite element analysis of a femoral surface replacement cup with Particular attention given to the interfacial stresses. An axisymmetric model with a grid of two-dimensional isoparametric elements is generated. This study utilizes an adaptive model so as to include the effect of bone remodeling in response to stress. The stiffness is considered to be linearly proportional to the stress in the femoral head and neck. Isotropic properties are chosen. The stresses are determined by performing iterative finite element analyses based upon an appropriate variation of the stiffness property in each analysis until the results obtained are within the convergence limit. Plots of stress contours in the femoral head are obtained for the loads acting on the joint during walking.